How potency is measured

Reversed-phase HPLC with UV detection, measured against a reference standard of documented identity and purity. Identity and purity are reported alongside, but potency is the determinative number.

Why potency, not purity, is the number that matters

Purity describes what fraction of the material present is the target compound. It clears specification in the substantial majority of samples. Potency describes how much of that compound is actually there. A vial can be 99% pure and still hold 28% less peptide than the label claims, because the remaining mass is counterion, residual water and excipient. Purity is about composition. Potency is about quantity.

The HPLC assay

Technique
Reversed-phase HPLC, UV detection
Instrument
Agilent 1100 Series HPLC with a G1365B multiple wavelength UV detector and a C18 reversed-phase column, controlled by ChemStation
Reference standards
Research-grade peptides from MilliporeSigma (Sigma-Aldrich), each supplied with the manufacturer’s certificate of analysis documenting identity and purity
Limit of detection
Determined for each peptide during method validation (signal-to-noise about 3:1)
Limit of quantitation
Determined for each peptide during method validation (signal-to-noise about 10:1)
Recovery
Acceptance criterion 98–102%
Repeatability
Acceptance criterion ≤2% RSD (CV)

The limits of detection and quantitation for your peptide are reported on your certificate of analysis, with the reference standard it was measured against.

Endotoxin testing

Bacterial endotoxin is quantified using a kinetic chromogenic assay (FireGene Kinetic Chromogenic Endotoxin Test Kit, TAL-based), read at 405 nm on a Thermo Scientific Multiskan FC microplate reader with temperature-controlled incubation at 37 °C. Each run includes a fresh standard curve prepared from the kit’s Control Standard Endotoxin (CSE), along with a positive product control (spike recovery).

Acceptance criteria are aligned with USP <85>: standard curve |r| ≥ 0.980 and spike recovery of 50–200%. Results are reported in EU/mL; values below the lowest standard are reported as “< [value] EU/mL.”

What the method does not tell you

Potency and purity describe the peptide. They say nothing about sterility, endotoxin load, or heavy metals. Endotoxin testing is offered as a separate assay on the potency, purity and endotoxin panel. Sterility and heavy-metal testing are not offered.